Changing Meter Setup Options
Setup options 1–6 and 10 have default settings (underlined in the table). To change the default
setting, use the table below. Settings for options 7–9 must be entered from the setup menu.
Setup
How To Get There
1
Turn the meter on and press
SETUP. Press ENTER.
2
Turn the meter on and press
SETUP
Press the up arrow once.
.
Press ENTER.
3
Turn the meter on and press
SETUP
Press the up arrow twice.
.
Press ENTER.
4
Turn the meter on and press
SETUP
Press the up arrow three
.
times. Press ENTER.
5
Turn the meter on and press
SETUP
Press the up arrow four
.
times. Press ENTER.
6
Turn the meter on and press
SETUP
Press the up arrow until
.
the setup number is
Press ENTER.
7
Turn the meter on and press
SETUP
Press the up arrow until
.
the setup number is
Press ENTER
8
Turn the meter on and press
SETUP
Press the up arrow until
.
the setup number is
ENTER
.
9
Turn the meter on and press
SETUP
Press the up arrow until
.
the setup number is
Press ENTER
10
Turn the meter on and press
SETUP
Press the up arrow until
.
the setup number is
Press ENTER
Laboratory pH/ISE Meter Quick Reference Card
Choose between using the BNC or 5-pin connector. ENTER
toggles between the options. Press EXIT when selection is
complete.
Turn display lock On or Off. ENTER toggles between the
options. Press EXIT when selection is complete.
Choose between °C or °F for displayed temperatures.
ENTER toggles between the options. Press EXIT when
selection is complete.
Select measurement resolution from tenths, hundredths, or
thousandths. ENTER toggles between the options. Press
EXIT when selection is complete.
Select pH 6.86 or 7.00 as the pH buffer the meter will
automatically recognize during calibration. ENTER toggles
between the options. Press EXIT when selection
is complete.
Enter the stability criteria (stability drift rate) using the
number keys (the default value is 0.500 mV). Press ENTER
6
.
to accept the value. Press EXIT when selection is complete.
Select the time of day by pressing the desired number key.
The meter uses a 24-hour clock. All four places must have a
7
.
value. Use a 0 if the left-most digit does not have a value
(i.e., 03:08). Press ENTER to accept the value. Press EXIT
.
when selection is complete.
Set the month and day by pressing the desired number
keys. The month uses the two places left of the slash; the
8
. Press
day of the month occupies the two places right of the slash.
Enter a 0 if the left-most digit does not have a value
(i.e., 02/06). Press ENTER to accept the value. Press EXIT
when selection is complete.
Set the year by using the number keys. Every place must
have a value. Press ENTER to accept the value. Press EXIT
9
.
when selection is complete.
.
Select the print interval by scrolling through the options with
the arrow keys (default is Off). When the desired option is
10
.
displayed, press ENTER to accept the value. Press EXIT
when selection is complete.
.
Description of Option
51775-23

Advertisement

loading
Need help?

Need help?

Do you have a question about the Sension4 and is the answer not in the manual?

Questions and answers

Summary of Contents for Hach Sension4

  • Page 1 51775-23 Laboratory pH/ISE Meter Quick Reference Card Changing Meter Setup Options Setup options 1–6 and 10 have default settings (underlined in the table). To change the default setting, use the table below. Settings for options 7–9 must be entered from the setup menu. Setup How To Get There Description of Option...
  • Page 2 8. After the last standard is measured, press will appear. Store ? EXIT 9. To save the calibration, press . To exit without saving the calibration, press ENTER EXIT © Hach Company, 1999. All rights reserved. Printed in the U.S.A hm/dk 5/99...