Welch Allyn MICROTYMP 4 User Manual page 47

Hand held portable tympanometer
Table of Contents

Advertisement

Welch Allyn MicroTymp® 4 Tympanometer User Manual
problem persists, contact Welch
Allyn
"WARNING! DEVICE UNCALIBRATED.
One or more default values require
recalibration before further tests are
performed
WARNING! DEFAULTS RELOADED.
Default configuration settings
reloaded. Check before making new
tests
WITHDRAW PROBE
Volume outside range
WITHDRAW PROBE
Blocked probe
WITHDRAW PROBE
INSERT PROBE
DIR 80024157 Ver A
This message should never normally be seen. If it
persists, contact your Welch Allyn service center.
This message should never be seen. Check all the
CONFIGURATION settings before taking any
measurements. If the error persists, contact your
Welch Allyn service center.
The probe has been moved during measurement. Re-
insert the probe to repeat the test.
The ear canal volume is above the 5ml. This message
also occurs when the probe is not properly inserted
into the ear.
The ear canal volume is below 0.1ml. This message
also occurs when the probe tip is blocked. Check that
the probe is correctly inserted into the ear. Check
that the probe is not blocked.
The seal was lost. Reinsert the probe to repeat the
test.
Page 34