Error Messages - Welch Allyn MICROTYMP 4 User Manual

Hand held portable tympanometer
Table of Contents

Advertisement

Welch Allyn MicroTymp® 4 Tympanometer User Manual
 PRINT (Print the results)
 SAVE RESULTS (Save the results in the internal database)
 VIEW TEST (Review the results as described above)
 MAIN MENU (Return to the main menu)
The results of the last test performed remain available even if the MicroTymp 4 has been turned
off. To view these results select VIEW THE LAST TEST from the main menu. After selecting the
required ear the tympanogram will be displayed. It will then be possible to view the results and
select the PROCESS RESULTS menu as if the test had just been completed.
NOTE: Results of the last test will be erased as soon as a new test is started. Test results
should be saved to the internal database or printed to ensure that data is not lost.
E
RROR MESSAGES
The following error messages may be seen during the test sequence.
Message Displayed
WITHDRAW PROBE
Volume outside range
WITHDRAW PROBE
Blocked ear
WITHDRAW PROBE
INSERT PROBE
DIR 80024157 Ver A
Indicator
Likely Cause(s)
Status
Yellow
The probe has been moved during measurement.
Flashing
Re-insert the probe to repeat the test.
Yellow
The ear canal volume is above the 5ml. This
message can also occur when the probe is not
Flashing
properly inserted into the ear.
Green
The ear canal volume is below 0.1ml. Check that
Flashing
the probe is correctly inserted into the ear. Also
check that the probe is not blocked.
Yellow
The seal was lost. Reinsert the probe to repeat
Flashing
the test.