Alert Messages And Display Information - Welch Allyn Micropaq 402 Service Manual

Table of Contents

Advertisement

Service Manual

Alert Messages and Display Information

Message and Display
Information
LOW BATTERY
VERY LOW BATTERY
BATTERY TOO LOW
SHUT DOWN IN PROGRESS
ACUITY CONNECTION LOST
EXCESSIVE ECG OFFSET
REPLACE ELECTRODES
Chest icon is displayed with
flashing ECG electrode(s).
NO ECG CABLE DETECTED
NO SpO2 SENSOR DETECTED
DEFECTIVE SpO
or
UNRECOGNIZED SpO
<key name> KEY STUCK
System Error
Thread: <nnn>
Error ID: <nnn>
a.
This alert message can be acknowledged from Acuity, but not from the monitor.
Possible Cause(s) and Suggested Response
The monitor will shut down in less than 30 minutes due to a low battery.
• Replace the battery as soon as possible.
The monitor will shut down in less than 5 minutes due to a low battery.
• Replace the battery as soon as possible.
The battery is so low the monitor has to shut down operation.
• Replace the battery immediately.
The monitor is not connected to the network.
• Press
disconnected from the network, the off-network icon and the yellow LED
continue to flash.
NOTE: The monitor will continue to attempt to reconnect until it is successful.
The monitor detects poor ECG electrode contact.
a
• Check and replace ECG electrodes as needed.
The monitor detects that at least one ECG electrodes is disconnected.
a
• Check and replace or reconnect electrodes as needed.
• If the ECG cable has been intentionally disconnected from the monitor, press
to cancel the alert tone.
• If the ECG cable has been unintentionally disconnected, reconnect it. Check the
patient and monitor to verify that ECG monitoring resumes properly.
The SpO
sensor has been disconnected for more than 5 seconds.
2
• If disconnection is intentional, press
• If disconnection is not intentional, reconnect the sensor or replace the sensor and
reconnect.
a
SENSOR
The SpO
sensor is either defective or not recognized.
2
2
• Replace the SpO
a
SENSOR
2
a
During the power-up self test, the monitor detected that a key is stuck (
or
). This can happen if you accidentally press a key down before the Main Menu
is displayed during power-up.
• Remove and then reinsert the battery to power up again and see if the key is still
stuck. If it is, contact your biomedical engineering department.
The equipment problem is so serious the monitor cannot be used.
• Contact your biomedical engineering department.
to acknowledge and silence the tone and cancel the message. While
to acknowledge and silence the tone.
sensor with a new, compatible SpO
2
Troubleshooting
25
sensor.
2
,
,
,