Error Message (Red-Backlit Screen And Normal Screen Are Displayed Alternately) - Panasonic Aicure ANUP5255V2 Operation Manual

Uv-curing system
Table of Contents

Advertisement

Aicure ANUP5255V2
Troubleshooting
14.2 Error message (Red-backlit screen and normal screen are
displayed alternately)
Message
UV intensity error
Cover open
Lamp illumination error
Overheat
Shutter error
Lamp-change indication
UV control
UV intensity
0000
Intensity rate 000 setting
Backlight
Green
(ON)
* The "Replace" is displayed in
reverse video at the top right-hand
corner of the display screen.
When the thermal relay is restored to normal condition after cooling, the error is reset.
Caution:
70
Error description
When UV control is ON and
the intensity control rate is
20% or more, the UV
intensity changes by ±5% of
the set value or more.
When the intensity control
rate is less than 20%, the UV
intensity changes by ±7% of
the set value or more.
The left side cover is
disconnected or the screws
are loose.
The lamp is not cooled
sufficiently.
The lamp was forcefully
turned off when the cover
was removed or for another
reason.
The temperature inside the
body rises abnormally high.
Irradiation shutter failure
The cumulative lamp OFF
ON Replace
time is approaching the
lamp-change time. (30 hours
0000
before)
When UV control is ON, the
intensity control rate is 98%
or more.
Remedy
(1) Conduct UV
intensity check.
(See Page 41.)
(2) Replace the lamp.
Attach the left side
cover.
After turning the "LAMP
ON/OFF" switch to OFF
and "POWER" switch to
OFF, turn the "POWER"
to ON again, and leave
until the cooling fan
stops (for 10 minutes).
With the "POWER"
switch ON, turn the
"LAMP ON/OFF" switch
to OFF and leave for 10
minutes. (Caution)
Contact the store from
which you made
purchase.
Replace the lamp. (See
Page 61.)
Use with a lower
intensity control rate.
If a necessary UV
intensity cannot be
obtained, the expiration
of the lamp's useful life
is assumed. Replace
the lamp.