Technical Specifications - Beurer BM 47 Instruction Manual

Table of Contents

Advertisement

Available languages

Available languages

T echnical alarm – description
Should the recorded blood pressure (systolic or diastolic) lie
outside the limits specified in the section "Technical speci-
fications", the technical alarm will appear on the display
indicating either " " or "Lo". In such cases, you should seek
medical assistance and check the accuracy of your proce-
dure.
The limit values for the technical alarm are factory set and
cannot be adjusted or deactivated. These alarm limit val-
ues are accorded second priority under the standard IEC
60601-1-8.
The technical alarm is a non-locking alarm and must not be
reset. The signal shown on the display will disappear auto-
matically after about 8 seconds.
NOTE:
If the unit still does not work, contact customer service at
1-800-536-0366.

10. Technical Specifications

Model no.
BM 47
Measurement
Oscillometric, non-invasive blood
method
pressure measurement on the up-
per arm
Measurement range Cuff pressure 0-300 mmHg,
Display accuracy
Measurement
inaccuracy
Memory
Dimensions
Weight
Cuff size
Permissible operat-
ing conditions
Permissible storage
conditions
Power supply
23
systolic 60-260 mmHg,
diastolic 40-199 mmHg,
Pulse 40-180 beats/minute
Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Max. permissible standard devia-
tion according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
4 x 30 memory spaces
5.3 in x 4 in x 2 in
(134 mm x 102 mm x 52.5 mm)
Approximately 280g
(without batteries and cuff)
8.7 in - 16.5 in (22 - 42 cm)
+50 °F to +104 °F (+10 °C to +40 °C),
≤ 85 % relative air humidity (non-
condensing)
-4 °F to +122 °F (-20 °C to +50 °C),
≤ 85 % relative air humidity, 800-
1050 hPa ambient pressure
4 x 1,5 V
AA batteries