Operating Conditions: Installation Instructions; Installation Instructions; Inlet And Outlet Run - Endress+Hauser Flowphant T DTT31 Technical Information

Flow switch for safe monitoring of mass flow and temperature in industrial processes
Hide thumbs Also See for Flowphant T DTT31:
Table of Contents

Advertisement

Installation instructions

8

Operating conditions: Installation instructions

• The sensor tip should be completely surrounded by medium.
• Position the sensor tip in the area of maximum fluid velocity
(pipe center).
• Minimum sensor immersion length: L
Orientation
• For horizontal pipes: lateral installation.
!
Note!
Installation from above only if the pipe is completely filled
with medium.
• For vertical pipes: installation in the ascending pipeline.
• DTT35: min. 3° inclination, because of self draining.
"
Caution!
• Do not install in down pipes open towards the end.
• The sensor tip should never touch the pipe wall.
• Electrical rotation of display by 180°
• Housing can be rotated up to 310 ° mechanically
Orientation
No restrictions, but self draining.

Inlet and outlet run

The sensor requires a fully developed flow profile for correct monitoring. For this reason, steadying sections
(5x DN) must be provided in the pipe after a pump, pipe bend, internal fittings and cross-sectional changes.
≥10 mm (0.4 in)
i
Installation conditions
Correct orientation
Incorrect installation!
DTT31, DTT35
A0007163
A0006977
A0006978
Endress+Hauser

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Flowphant t dtt35

Table of Contents