Fluke Biomedical ESA614 Getting Started Manual page 27

Electrical safety analyzer
Table of Contents

Advertisement

Patient Load ........................................................ AAMI ES1-1993 Fig. 1
Crest factor .......................................................... ≤3
Ranges ................................................................ 0.0 μA to 199.9 μA
Accuracy
DC to 1 kHz ..................................................... ±(1 % of reading + (1 μA or 1 LSD, whichever is greater))
1 kHz to 100 kHz ............................................. ±(2 % of reading + (1 μA or 1 LSD, whichever is greater))
1 kHz to 5 kHz (current > 1.6 mA) ................... ±(4 % of reading + (1 μA or 1 LSD, whichever is greater))
100 kHz to 1 MHz ............................................ ±(5 % of reading + (1 μA or 1 LSD, whichever is greater))
Accuracy for Isolation and MAP leakage tests all ranges are:
At 120 V ac + (2.5
At 230 V ac additional
Mains on applied part test voltage ...................... 100 % ±7 % of Mains for AAMI, current limited to 1 mA ±25 % per AAMI
Insulation resistance
Ranges ................................................................ 0.5 to 20.0 MΩ
Accuracy
20 MΩ Range .................................................. ±(2 % of reading + 0.2 MΩ)
100 MΩ Range ................................................ ±(7.5 % of reading + 0.2 MΩ)
Source test voltage ............................................. 500 or 250 V dc (+20 %, -0 %) 2.0 ±0.25 mA short-circuit current
Maximum load capacitance ................................. 1 μF
200 μA to 1999 μA
2.00 mA to 10.00 mA
μ
A or 1 LSD, whichever is greater)
±
μ
3.0 % and + (2.5
A or 1 LSD, whichever is greater)
20.0 to 100.0 MΩ
Note
Electrical Safety Analyzer
Detailed Specifications
21