Fluke Biomedical ESA614 Getting Started Manual page 28

Electrical safety analyzer
Table of Contents

Advertisement

ESA614
Getting Started Manual
ECG Performance Waveforms
Accuracy
Frequency ....................................................... ±2 %
Amplitude ........................................................ ±5 % of 2 Hz square wave only, fixed @ 1 mV Lead II configuration
Waveforms
ECG Complex ................................................. 30, 60, 120, 180, and 240 BPM
Ventricular Fibrillation
Square wave (50 % duty cycle) ...................... 0.125 Hz and 2 Hz
Sine wave ....................................................... 10, 40, 50, 60, and 100 Hz
Triangle wave .................................................. 2 Hz
Pulse (63 ms pulse width) ............................... 30 BPM and 60 BPM
Factory-Supplied Test Sequences
NFPA99 Patient Monitor
NFPA99 Defibrillator
NFPA99 Infusion Pump
NFPA99 Ultrasound Device
NFPA99 Generic Device
ANSI/AAMI ES-1 Patient Monitor
ANSI/AAMI ES-1 Defibrillator
ANSI/AAMI ES-1 Infusion Pump
ANSI/AAMI ES-1 Ultrasound Device
ANSI/AAMI ES-1 Generic Device
22