FIGURE 7.1 Cleaning the Cabinet Filter
Cleaning the Cabinet
1. Clean the cabinet with a mild household cleaner and
non-abrasive cloth or sponge.
Humidifier Cleaning and Thermic Disinfection
NOTE: Clean and disinfect the oxygen humidifier daily to reduce
limestone deposits and eliminate possible bacterial contamination.
Follow the instructions provided by the manufacturer. If none are
provided, follow these STEPS:
1. Wash humidifier in soapy water and rinse with a solution
of ten parts water and one part vinegar.
2. Rinse thoroughly with hot water.
3. Immerse humidifier completely in water at 65°C for 15
minutes.
4. Air dry thoroughly.
NOTE: To limit bacterial growth, air dry the humidifier thoroughly
after cleaning when not in use. Refer to Set Up on page 17 for use.