SECTION 5—TYPICAL PRODUCT PARAMETERS
Flow Range:
Potential Obstruction
Alert:
Power
Consumption:
Pressure Relief
Mechanism
Operational at:
Change in maximum
recommended flow
when back pressure of
7kPa is applied:
Filters:
Safety System:
Width:
Height:
Depth:
Weight:
Shipping Weight:
Operating Ambient
Conditions:
Exhaust Cooling Air
Temperature:
Oxygen Output
Temperature:
Cabinet:
No AP/APG
Standards and
Regulatory Listing:
CE marked model
Perfecto
™Series
2
0.5 to 5 L/min (maximum)
For flowrates less than 1 L/min, we recommend the use of the
Invacare Pediatric Flowmeter Accessory (IRCPF16AW)
0 L/min to 0.5 L/min
The concentrator detects a condition that may indicate a potential
obstruction of the output oxygen.
Rapid audible beeping alert (this alert is deactivated when
accessories are connected). May be associated with flow setting of
0.5 L/min or less.
Perfecto
AW - 300 W @ 5 L/min, 280 W @3 L/min
2
Perfecto
VAW - 320 W@ 5 L/min
2
241 kPa ± 24.1 kPa (35 psi ± 3.5 psi)
0.7 L/min
Cabinet, Outlet HEPA and Compressor Inlet
Current overload or line surge shutdown
High temperature compressor shutdown
High Pressure Alarm w/compressor shutdown Low Pressure Alarm
w/compressor shutdown Battery Free Power Loss Alarm
SensO2 Oxygen System Possible Obstruction Alert.
38.1 cm ± 1 cm (15 in ± 3/8 in)
58.4 cm ± 1 cm (23 in ± 3/8 in)
30.5 cm ± 1 cm (12 in ± 3/8 in)
Perfecto
AW Models - 20.5 kg ± 1 kg (45 lbs ± 2 lbs)
2
Perfecto
VAW Models - 18.1 kg ± 1 kg (40 lbs ± 2 lbs)
2
Perfecto
AW Models - 22.7 kg ± 1 kg (50 lbs ± 2 lbs)
2
Perfecto
VAW Models - 20.4 kg ± 1 kg (45 lbs ± 2 lbs)
2
10°C - 35°C (50°F - 95°F) at 20-60% relative humidity
Less than Ambient +19°C (+ 45°F)
Less than Ambient +3°C (+ 8°F)
Impact Resistant flame-retardant plastic cabinet that conforms to UL
94-V0
Not suitable for use in the presence of a flammable anaesthetic
mixture
IEC/EN 60601-1, A1, A2
IEC/EN 60601-1-2
IEC/EN61000-3-2,
IEC/EN61000-3-3
ISO8359
MDD 93/42/EEC, Annex I and IX
IRC5PO2AW & IRC5PO2VAW
14
Part No 1163147

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Perfecto2 irc5po2awPerfecto2 irc5po2vaw

Table of Contents