7. Select the location for the humidifier bottle/adapter
assembly:
•When placing the humidifier bottle in the
compartment on the concentrator, attach the
humidifier bottle/adapter tube to the oxygen outlet
connector on the concentrator. See Detail "B" in
FIGURE 6.3.
•To position the humidifier bottle in a patient side
location, place the humidifier bottle/adapter
assembly in the humidifier bottle holder (M1521,
Refer to Optional Accessories on page 31) and place
on a flat surface as close to the patient as possible.
Attach an oxygen tubing connector (MS4301, Refer
to Optional Accessories on page 31) to the end of
the humidifier bottle/adapter tube. Connect one
end of the oxygen supply tubing (MS4107 or
MS4121, Refer to Optional Accessories on page 31)
to the free end of the oxygen tubing connector and
the opposite end to the oxygen outlet connector on
the concentrator.
NOTE: When using tubing connections longer than 2.1 m (7 feet)
in length, position the humidifier as close to the patient as possible
to allow for maximum humidification output.
8. Attach the patients' nasal cannula supply tube to the
humidifier bottle outlet. See Detail "B" in FIGURE 6.3.
9. After assembly, ensure that oxygen is flowing through the
nasal cannula.
Part No 1163147