Electrical:
Placement:
Tubing:
Time of Operation:
Recommended
Storage and Shipping
Temperature:
Environment:
Minimum Operating
Time:
LABEL
SYMBOL
O
2
Part No 1163147
SECTION 5—TYPICAL PRODUCT PARAMETERS
No extension cords
No closer than 30,5 cm (12 in) from any wall, furniture, draperies, or
furniture to assure sufficient air flow
Avoid deep pile carpets and heaters, radiators or hot air registers
Floor location only
No confined spaces (Example: No closets)
2 m (7 ft) cannula with a maximum 15 m (50 ft) of Crush-Proof
Tubing (DO NOT pinch)
Up to 24 hours per day
-29°C to 65°C (-20°F to 150°F) at 15-95% relative humidity
Smoke and soot-free
No confined spaces (Example: No closets)
30 Minutes
O
INDICATORS
2
O
PURITY
2
SYSTEM OKAY
O
over 85%
2
O
Between 73%
2
to 85%
SYSTEM FAILURE
O
Below 73%
2
INDICATOR LIGHTS (LED)
GREEN Indicator Light
YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor
Failure
Call a qualified technician.
RED Indicator Light
Continuous Audible Alarm
Sieve-GARD Compressor Shutdown
15
Perfecto
™Series
2