Product Disposal and Recycling ..............9 Product Support ..................9 Introduction Thank you for purchasing your R6001 Thermo-Hygrometer. Please read the following instructions carefully before using your instrument. By following the steps outlined in this manual your meter will provide years of reliable service.
This product has been manufactured in an ISO9001 facility and has been calibrated during the manufacturing process to meet stated product specifications. If a certificate of calibration is required please contact the nearest authorized REED distributor or authorized Service Center. Please note an additional fee for this service will apply. Safety Never attempt to repair or modify your instrument.
Operating Humidity Range: 10 to 80% Dimensions: 9.5 x 2.1 x 1.3" (240 x 54 x 34mm) Weight: 6.4oz (180g) Included • Carrying Case • Battery Instrument Description 1. Sensor Protective Cap 2. LCD Display 3. Power Button 4. MAX/MIN Button 5. Data Hold Button 6. Relative Function Button 7. Unit of Measure Button 8. Battery Cover REED-Direct.ca www.REEDINSTRUMENTS.com info@REED-Direct.ca 1.800.561.8187...
Note: When the Data Hold function is enabled the REL, MAX/MIN, and °C/°F buttons are disabled. Relative Function Press the REL button to store the current measurement. The meter will display the difference between any new measurement and the stored measurement. Press the REL button again to exit Relative Function mode and resume normal operation. continued... REED-Direct.ca www.REEDINSTRUMENTS.com info@REED-Direct.ca 1.800.561.8187...
If the meter is turned off then back on, the "Auto Power Off" feature will be enabled again. Humidity Calibration Procedure In order to calibrate the meter humidity calibration standards are required (REED R9933, R9975 or R9980). If the meter has been exposed to extreme conditions or it has not been used in a long period of time the meter should be calibrated by following steps 1 through 6 below. Turn the meter off.
When the LCD flashes, press the REL button to enter calibration mode. This will be indicated by "CAL 1" on the LCD. Insert the tip of the meter into the 33%RH Humidity Calibration Standard (REED R9933) then press the MAX/MIN button to start the adjustment. The calibration adjustment will be confirmed by a long beep. The meter will then display "CAL 2" to calibrate the second calibration point. Insert the tip of the meter into the 75%RH Humidity Calibration Standard (REED R9975) then press MAX/MIN button to start the adjust- ment. The calibration adjustment will be confirmed by a long beep.
REED Instruments guarantees this instrument to be free of defects in material or workmanship for a period of one (1) year from date of shipment. During the warranty period, REED Instruments will repair or replace, at no charge, products or parts of a product that proves to be defective because of improper material or workmanship, under normal use and maintenance.
Product Support If you have any questions on your product, please contact your authorized REED distributor or REED Instruments Customer Service by phone at 1-877-849-2127 or by email at info@reedinstruments.com. Please visit www.REEDINSTRUMENTS.com for the most up-to-date manuals, datasheets, product guides and software.