Step Inputs

Logic
Analyzer
w/ EyeScan
Driver
Rsource
50 Ω
E5400-Pro Series Soft Touch User's Guide
Operating the E5402A, E5405A, and E5406A-Pro Series Probes
Maintaining signal fidelity to the logic analyzer is critical if
the analyzer is to accurately capture data. One measure of a
system's signal fidelity is to compare V
step inputs. For the following graphs, V
logic analyzer probe tip. Eye Scan was used to measure V
the signal seen by the logic analyzer. The measurements
were made on a mid- bus connection to a 50Ω transmission
line load terminated at the receiver. These measurements
show the logic analyzer's response while using the E5402A,
E5405A, and E5406A- pro series soft touch probes.
E5382A
Probe
Step
output
Z0=50 Ω
Figure 35
E5406A)
The following plots were made on an Agilent 54750A
oscilloscope and an Agilent 16760A logic analyzer using an
Agilent 8133A pulse generator with various rise time
converters.
54701A
Probe
Z0=50 Ω
Step input measurement schematic (E5402A, E5405A, and
to V
for various
in
out
is the signal at the
in
out
Oscilloscope
2.5GHz BW
incl. probe
Receiver
Rterm
50 Ω
4
,