ACTIVATE A NEW POD

+
Assemble the following supplies:
DASH
Personal Diabetes Manager (PDM)
Sealed DASH
Pod, indicated by a blue needle cap
Vial of U-100, rapid acting insulin at room temperature.
(See the Omnipod DASH
insulins tested and found to be safe with the Omnipod
DASH
Insulin Management System)
Alcohol prep swab
+
Wash your hands
1.
+
To set up a new Pod,
Tap "SET UP NEW
POD"
+
Read and perform each
instruction carefully
WARNING:
+ NEVER inject air into the fill port. Doing so may result in unintended or interrupted insulin delivery
+ NEVER use a Pod if you hear a crackling noise or feel resistance when you depress the plunger.
These conditions can result in interrupted insulin delivery
CAUTION:
Do not use any other type of needle or filling device besides the fill syringe provided with each Pod.
OMNIPOD DASH
System User Guide for
SYSTEM INSTRUCTIONS |
Activate a New Pod