OMNIPOD
SYSTEM INSTRUCTIONS |
®
ACTIVATE A NEW POD
+
Assemble the following supplies:
–
Vial of insulin at room temperature
(U-100, rapid-acting). See the Omnipod
Insulin Management System User Guide for
the insulins that are recommended for use
with the Omnipod
®
–
One sealed Pod
–
PDM
–
Alcohol prep swab
+
Wash your hands.
1. Fill the Pod
1.1
+
Remove the Pod from its
sterile packaging.
+
Use the alcohol prep swab
to clean the top of the
insulin vial.
+
Assemble the fill syringe
by twisting the needle onto
the syringe.
WARNING:
+ NEVER inject air into the fill port. Doing so may result in unintended or interrupted insulin delivery
+ NEVER use a Pod if you hear a crackling noise or feel resistance when you depress the plunger.
Activate a new Pod
®
System
1.2
+
Remove the protective cap.
1.3
+
Draw air into the fill syringe
equal to the amount of insulin
indicated in your Diabetes
Management Plan.
+
Depress air into the vial
of insulin.
+
Turn the vial and syringe
upside down.
+
Withdraw insulin from the
vial and fill the syringe
with the amount of insulin
determined by your healthcare
professional; fill it at least to
the MIN line.
+
Remove any air bubbles
from the syringe.
CAUTION:
Do not use any other type of needle or filling device besides
the fill syringe provided with each Pod.
Need help?
Do you have a question about the PODDER and is the answer not in the manual?
Questions and answers