Maximum wiring distances
The maximum distance between the sensor and transmitter depends on cable type. See Table 2.
Table 2
Maximum cable lengths
Cable type
Micro Motion 9-wire to an MVD
transmitter or core processor
Micro Motion 9-wire to all other
transmitters
Micro Motion 4-wire
User-supplied 4-wire
Power wires (VDC)
Signal wires (RS-485)
Pipe run
Micro Motion sensors do not require a straight run of pipe upstream or downstream.
Valves
After the sensor and transmitter have been installed, you must perform the zeroing procedure. During the
zeroing procedure, flow through the sensor must be halted and the sensor tubes must be completely full of
process fluid. A shutoff valve, downstream from the sensor, is recommended to halt flow during the zeroing
procedure. For more information about zeroing, refer to the instruction manual shipped with the transmitter.
Sensor Installation: F-Series
Wire size
Not applicable
Not applicable
Not applicable
2
22 AWG (0,35 mm
)
2
20 AWG (0,5 mm
)
2
18 AWG (0,8 mm
)
2
22 AWG (0,35 mm
) or larger
Determining a Location
Maximum length
60 feet (20 meters)
1000 feet (300 meters)
1000 feet (300 meters)
300 feet (90 meters)
500 feet (150 meters)
1000 feet (300 meters)
1000 feet (300 meters)