E . Device Specifications
Model Name: CU-SP1 AUTO
Physical
Category
Dimensions
Weight
Environmental
Category
Operational Status (The device is in emergency use)
Standby Status (The device is stored together with the defibrillator pads and the battery pack
is inserted - ready to be used in an emergency)
Transport Status (device only, no defibrillator pads and battery pack included)
Altitude
Drop
Vibration
Sealing
ESD
EMI (Radiated)
EMI (Immunity)
Nominal Specifications
260㎜ x 256㎜ x 69.5㎜ (Width x Length x Height)
2.4㎏ (Including the battery pack and pads)
Nominal Specifications
Temperature: 0℃ ~ 43℃ (32℉ ~ 109℉)
Humidity: 5% ~ 95% (non condensing)
Temperature: 0℃ ~ 43℃ (32℉ ~ 109℉)
Humidity: 5% ~ 95% (non condensing)
Temperature: -20℃ ~ 60℃ (-4℉ ~ 140℉)
Humidity: 5% ~ 95% (non condensing)
0 to 15,000 feet (operational and storage)
Withstands 1.2-meter drop to any edge, corner, or surface
Operating: Meets MIL-STD-810G Fig.514.6E-1, random
Standby: Meets MIL-STD-810G Fig.514.6E-2, swept sine(helicopter)
IEC 60529: IP55
Meets IEC 61000-4-2:2001
Meets IEC 60601-1-2 limits, method EN 55011:2007 +A2:2007,
Group 1, Class B
Meets IEC 60601-1-2 limits, method EN 61000-4-3:2006 +A1:2008 Level 3
(10V/m 80MHz to 2500MHz)
61
Need help?
Do you have a question about the i-PAD CU-SP1 AUTO and is the answer not in the manual?
Questions and answers
How many bars should be showing on thew battery when its fully charged. It has been showing 2 bars for 3 months. The green light is flashing!!