4.2 Preparation for Defibrillation
① Set the Adult/Pediatric Selection Switch to match the victim.
Adult victim
• Open the switch cover
• Set the switch to adult defibrillation mode as shown in the following picture
Child victim (victim is under 25kg or 8 years old)
If the pediatric pads are attached, the i-PAD CU-SP1 AUTO automatically adjusts its
defibrillation energy output for pediatric defibrillation regardless of the position of the
Adult/Pediatric Selection Switch (i.e. the output will be pediatric even if the selection switch is
set to adult)
If there are no pediatric pads for the pediatric patient, adult pads may be used. Ensure that the
Adult/Pediatric Selection Switch is set to Pediatric Mode. If the switch has not been set yet,
move it to Pediatric Mode as shown in the figures below
• Open the switch cover
17