CU Medical Systems i-PAD NF1200 Operator's Manual

CU Medical Systems i-PAD NF1200 Operator's Manual

Intelligent public access defibrillator
Hide thumbs Also See for i-PAD NF1200:

Advertisement

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the i-PAD NF1200 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for CU Medical Systems i-PAD NF1200

  • Page 4 Edition 1 Notice i-PAD Operator’s Manual CU Medical Systems, Inc. reserves the right to make changes on the device specifications contained in this manual at any time without prior notice or obligation to customers. Printed in the Republic of Korea Publication Date: February 2006 Operator’s Manual Part No.: OMP-NF1-E-01...
  • Page 7 A.1 Standard Accessories A.2 Optional Accessories A.3 Service Center B. Equipment Symbols B.1 i-PAD Defibrillator B.2 i-PAD Packaging B.3 Symbols on Accessories B.3.1 Battery [CUSA0601F] B.3.2 Pads [CUA0512F] C. Glossary D. i-PAD NF1200 Specifications E. Electromagnetic Compatibility NF1200 Manual Edition 1...
  • Page 35 Dilute hydrogen peroxide Do not immerse any part of the i-PAD in fluids. If the i-PAD has been submerged in water, call CU Medical Systems, Inc. or an authorized representative for service assistance Do not allow fluid to enter the case of the device.
  • Page 45 Information Contact no. Brief description of the problem How to Contact Us CU Medical Systems, Inc. Medical Instrument Industry Park 1720-26, Taejang-dong, Wonju-si, Gangwon-do, Korea TEL : +82 33 747 7657 FAX : +82 33 747 7659 Homepage : www.cu911.com Sales inquiries : sales@cu911.com...
  • Page 46 - PC S/W (CU Expert Version 3.00) - i-PAD Trainer A.3 Service Center Customer Service Homepage www.cu911.com Address CU Medical Systems, Inc. Medical Instrument Industry Park 1720-26, Taejang-dong, Wonju-si, Gangwon-do, Korea +82 33 747 7657 +82 33 747 7659 Sales Inquiries E-mail sales@cu911.com...
  • Page 54 i-PAD NF1200 Specifications Model: NF1200 Physical Nominal Specifications Category 3.19 in high X 8.66 in wide X 11.06 in deep Size (81 mm high X 220 mm wide X 281 mm deep) - without handle and battery pack Weight Approximately 4.84 lbs (2.2 kg) with battery pack installed Environmental Category Nominal Specifications...
  • Page 56 NF1200 Specifications Waveform Specifications (200 Joules) Patient Impedance Phase A, Duration Phase B, Duration Energy Delivered (Ohms) (milliseconds) (milliseconds) (Joules) 11.0 11.0 12.8 12.8 NF1200 Manual Edition 1...
  • Page 61 . Electromagnetic Compatibility Guidance and manufacturer’s declaration - electromagnetic emissions The i-PAD is intended for use in the electromagnetic environment specified below. The customer or the user of the i-PAD should assure that it is used in such an environment. Emissions Test Compliance Electromagnetic environment - guidance...
  • Page 62 . Electromagnetic Compatibility Guidance and manufacturer’s declaration - electromagnetic immunity The i-PAD is intended for use in the electromagnetic environment specified below. The customer or the user of the i-PAD should assure that it is used in such an environment. Immunity Test IEC 60601-1 test Compliance level...
  • Page 63 Guidance and manufacturer’s declaration - electromagnetic immunity The i-PAD is intended for use in the electromagnetic environment specified below. The customer or the user of the i-PAD should assure that it is used in such an environment. Immunity IEC 60601-1 test Compliance Electromagnetic environment - Test...

Table of Contents