5.2 Saving and Transferring Treatment Data
5.2.1 Device Usage
This device automatically saves the following treatment data:
• ECG data
• Usage information
The treatment data is automatically recorded in the internal memory. This data is not erased
even if the device is turned off. The recorded treatment data may be transferred to a personal
computer (PC).
CAUTION
This i-PAD CU-SP1 AUTO keeps the data of the 5 most recent treatment operations and can
save up to 3 hours of ECG data for each rescue operation. ECG data beyond 3 hours will not
be recorded.
When the device is used more than 5 times, it deletes the oldest treatment data to make room
for data from a new treatment operation. It is recommended to transfer treatment data to a PC
after each use of the device.
If the battery pack is removed while the device is operating, treatment data cannot be properly
recorded. If you wish to remove the battery pack, turn the power off by pressing the Power
Button for at least 1 second before removing the battery pack.
5.2.2 Transferring Treatment Data
The treatment data may be transferred via a SD card or IrDA. The entire treatment data of all
patients that is recorded on the device is transferred using only the SD card method, while the
treatment data of one patient can be transferred using only the IrDA method.
1. Copying Treatment Data by Using the SD Card
① Format the SD card on the PC to FAT (FAT16) format.
② Open the SD Card Cover on the device and insert the SD card into the port.