Dear customer, Thank you for choosing a product from the Wassermann range. Wassermann Dental- Maschinen incorporates the highest standards of quality and the latest technology. In order to enjoy maximum performance and years of trouble-free operation, please read this user manual carefully before you connect this device and start work, and operate the device according to the recommended guidelines.
A quality pressure gauge which shows the pressure achieved with great accuracy, is recessed into the housing and is thus protected from dirt. The WW-33’s oxidation-free piston (plastic), tensile steel construction (no risk of breaking), and sturdy plastic-coated housing all contribute to the unit’s long service life.
Safety Symbols used in this Manual Warning! This is a warning of risk situations and dangers. Failure to observe this warning could be life-threatening. These warnings has to be observed. Information! This symbol draws your attention to specific features that has to be observed. Safety Guidelines Configuring and operating this equipment requires precise knowledge and observance of the instructions in this user manual.
Wassermann Dental-Maschinen GmbH are not extended by these instructions. Application With a continuous working pressure of up to 6 tonnes, the WW-33 hydraulic press is used to press caps individually and in accordance with the material used. Only use the device for this type of application.
Installation Open the box, remove the packing materials, and carefully lift out the device and accessories. Check the included accessories. The device has to stand horizontally on a steady and even surface. Install the device in a place where it will not block the working area and the functionality (take the dimensions into account).
Operation General Operating Instructions All instructions for using the unit, whether in verbal or written form, are based on our own experience and experimentation and can only be regarded as guidelines. 1. Prepare cap as per Point 7. 2. Turn the pressure valve (black knob on housing) to CLOSED. 3.
Troubleshooting Fault Cause Solution Little or no pressure Oil level below minimum and Vent press as explained below builds air bubble has entered Point 9.1 system during filling Pressure drops to Dirty pressure valve Pump strongly with newly filled oil zero despite enclosed (recommended oil) cap or piston rises...
Care and Maintenance 10.1 Cleaning The identification plate has always to be kept in easily legible condition and has not to be removed. Remove external dirt from time to time with some form of cold cleaner. Use only cold cleaners to avoid damaging the paintwork or removing the lettering. The equipment should be cleaned at regular intervals to ensure trouble-free operation.
Technical Data WW-33, 2–4 flasks Item no.: 126998 WW-33, 1–3 flasks Item no.: 126999* W x D / D with pump lever 200 x 158 / 295 mm Height max. 710 / min. 540 mm max. 650 / min. 480 mm*...
Dental-Maschinen GmbH for dental Rudorffweg 15-17 applications 21031 Hamburg Germany Model: WW-33 for 2-4 flasks (special design) Item no.126998 WW-33 for 1-3 flasks Item no.126999 Applicable DIN EN ISO 12100 standards: Hiermit wird bestätigt, dass die oben bezeichnete Maschine den genannten EU-Richtlinien entspricht.