Sense[N]:<Function>:Average[:State] - Keithley 2280 Series Reference Manual

Precision measurement dc power supplies
Hide thumbs Also See for 2280 Series:
Table of Contents

Advertisement

Section 7: SCPI command reference
:SENSe[n]:<function>:AVERage[:STATe]
This command enables or disables the averaging filter for measurements of the selected function.
Type
Command and query
Usage
:SENSe[n]:<function>:AVERage[:STATe] <state>
:SENSe[n]:<function>:AVERage[:STATe]?
[n]
<function>
<state>
Details
This command enables or disables the averaging filter. When this is enabled, the reading returned by
the instrument is an averaged value, taken from multiple measurements. The settings of the filter
count and filter type for the selected measure function determines how the reading is averaged.
Example
SENS:CONC:AVER:WIND 0.1
Also see
:SENSe[n]:<function>:AVERage:COUNt
:SENSe[n]:<function>:AVERage:TCONtrol
:SENSe[n]:<function>:AVERage:WINDow
7-62
Affected by
Recall settings
Instrument reset
Power cycle
Channel number: For single-channel instruments, the value is 1
The measurement function to which this setting applies:
Concurrent: CONCurrent[:DC]
Current: CURRent[:DC]
Voltage: VOLTage[:DC]
The filter status; set to one of the following values:
Disable the averaging filter: OFF or 0
Enable the averaging filter: ON or 1
(on page 7-61)
(on page 7-63)
(on page 7-64)
Series 2280 Precision DC Power Supply Reference Manual
Where saved
Save settings
Set the average filter window as 0.1%.
Default value
1 (ON)
077085503 / March 2019