Section 7: SCPI command reference

*LANG?

This command determines which command set is used by the instrument.
Type
Query only
Usage
*LANG?
Details
The only available remote command set is SCPI, which is an instrument-specific language built on
the SCPI standard.
Example
*LANG?
Also see
None

*RCL

This command returns the instrument to the setup that was saved with the *SAV command.
Type
Command only
Usage
*RCL <n>
[n]
Details
Restores the state of the instrument from a copy of its settings that are stored in the setup memory.
The settings are saved using the *SAV command.
Example
*RCL 3
Also see
*SAV
(on page 7-7)
Saving setups
7-6
Affected by
Not applicable
Affected by
Not applicable
An integer from 1 to 5 that represents the saved setup
(on page 2-82)
Series 2280 Precision DC Power Supply Reference Manual
Where saved
Nonvolatile memory
Verify the setting by sending the language query.
Output:
SCPI
The SCPI command set is in use.
Where saved
Not applicable
Restores the settings stored in memory location
3.
Default value
SCPI
Default value
Not applicable
077085503 / March 2019