Measurement system uncertainties
In accordance with ISO 80601-2-70:2015 the measurement uncertainty of the manufacturer's test equipment is:
For measures of flow
For measures of static pressure
For measures of dynamic pressure
For measures of volume (< 100 mL)
For measures of volume (≥ 100 mL)
For measures of time
Note: ISO 80601-2-70:2015 stated accuracies and test results provided in this manual for these items already include the
relevant measurement uncertainty from the table above.
Sound
Declared dual-number noise emission values in accordance with ISO 4871:1996
Pressure level measured according to ISO 80601-2-70:2015
(CPAP mode)
Power level measured according to ISO 80601-2-70:2015
(CPAP mode)
The A-weighted sound pressure level measured according to ISO 17510:2015:
Air tubing connected to AirMini F20 connector with mask
Air tubing connected to AirMini N20 or P10 connectors
with mask
The A-weighted sound power level measured according to ISO 17510:2015:
Air tubing connected to AirMini F20 connector with mask
Air tubing connected to the AirMini N20 or P10 connectors
with mask
Pressure flow curve (ISO 17510:2015)
Resistance
Inspiratory and expiratory resistance with AAV open to
atmosphere (ISO 17510:2015)
Inspiration at 50L/min
Expiration at 50L/min
AAV Pressures (ISO 17510:2015)
Open-to-atmosphere pressure
Closed-to-atmosphere pressure
Design life
Machine, power supply unit:
Air tubing, connectors:
HumidX:
General
The patient is an intended operator.
28
± 1.5 L/min or ± 2.7% of reading (whichever is greater)
± 0.15 cm H
O (hPa)
2
± 0.27 cm H
O (hPa)
2
± 5 mL or 6% of reading (whichever is greater)
± 20 mL or 3% of reading (whichever is greater)
± 10 ms
29 dBA with uncertainty of 2 dBA
37 dBA with uncertainty of 2 dBA
19 dBA with uncertainty of 3 dBA
15 dBA with uncertainty of 3 dBA
27 dBA with uncertainty of 3 dBA
23 dBA with uncertainty of 3 dBA
Pressure
[cm H
O (hPa)]
2
4
8
12
16
20
F20 connector
0.7 cm H
O (hPa)
2
0.5 cm H
O (hPa)
2
F20 connector
0.9 cm H
O (hPa)
2
1.0 cm H
O (hPa)
2
5 years
6 months
30 days
Flow [L/min]
21
26
29
31
33

Hide quick links: