Operating humidity:
Operating altitude:
Storage and transport temperature:
Storage and transport humidity:
Air filter
Standard:
Hypoallergenic:
AirMini machine
Dimensions:
Weight:
Housing construction:
Air outlet:
AirMini air tubing
Material:
Length:
Inner diameter:
Electromagnetic compatibility
The AirMini complies with all applicable electromagnetic compatibility requirements (EMC) according to
IEC60601-1-2:2014, for residential, commercial and light industry environments. Portable and mobile RF communications
equipment should be used no closer to any part of the machine, including cables, than the recommended 10 cm separation
distance. The AirMini has been designed to meet EMC standards. However, should you suspect that the machine
performance (eg, pressure or flow) is affected by other equipment, move the machine away from the possible cause of
interference.
The AirMini complies with Part 15 of the FCC Rules and Industry Canada license-exempt RSS standards. Operation is
subject to the following two conditions: This machine may not cause harmful interference, and this machine must accept
any interference received, including interference that may cause undesired operation.
FCC ID: QOQBT121, IC: 5123A-BGTBT121
Additional information regarding the FCC Rules and IC compliance for this machine can be found on
www.resmed.com/downloads/devices.
Declaration of Conformity (DoC to the Radio Equipment Directive)
ResMed declares that the AirMini device (models 381xx) is in compliance with the essential requirements and other
relevant provisions of Directive 2014/53/EU (RED). A copy of the Declaration of Conformity (DoC) can be found on
Resmed.com/productsupport
This radio equipment operates with the following frequency bands and maximum radio-frequency power: Bluetooth Class
2. 2402 to 2480 MHz, 4dBM.
Aircraft use
ResMed confirms that machine meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21,
category M) for all phases of air travel.
Wireless technology
Technology used:
Connection types:
Frequency:
Max RF power output:
Operating range:
26
10 to 95% relative humidity, non-condensing
Sea level to 2,591 m; air pressure range 1013 hPa to
738 hPa
-25°C to +70°C
5 to 95% relative humidity, non-condensing
Material: Polyester non woven fibre
Average arrestance: >75% when tested to EN779
Material: Blended synthetic fibres in a polypropylene carrier
Efficiency: >80% (average) when tested to EN 13274-7
136 mm (W) x 84 mm (D) x 52 mm (H)
300 g
Flame retardant engineering thermoplastic
Proprietary connector, 16mm inner diameter. Not compatible
with EN ISO 5356-1 connectors.
Flexible plastic
1.9 m
15 mm
Bluetooth
SPP, iAP2, GATT
2402 to 2480 MHz
+4 dBm
10 m Class 2