AirSense 10 and AirCurve 10 reprocessing guide
This guide is intended for multipatient use of the AirSense™10 and AirCurve™10 devices.
When the device is used for multiple patients, for example, in a sleep lab, clinic, hospital or at a health care provider, the
cleanable water tub, air outlet and air tubing should be reprocessed between each patient use.
If the cleanable water tub or the air tubing are being used for a single user in the home, refer to the cleaning instructions in this
guide or in the User Guide.
Described here are ResMed's recommended and validated procedures for cleaning and disinfecting the cleanable water tub, air
outlet and air tubing. However, the steps for disinfection vary regionally and each healthcare facility should consult its own
procedures before carrying out those within this guide.
Note: The standard water tub cannot be disinfected. If contaminated, it must be discarded and replaced with a new water tub.
WARNING
• ResMed cannot give any assurance that deviations from the procedures listed in this guide, and their effect on the
performance of the product, will be acceptable.
• When using detergents, disinfectants or sterilization agents, always follow the manufacturer's instructions.
• Beware of electrocution. Do not immerse the device, power supply or power cord in water. If liquids are spilled into
or onto the device, unplug the device and let the parts dry. Always unplug the device before cleaning and make sure
that all parts are dry before plugging it back in.
Surface disinfection
1. Wipe the exterior of the device including display, externally accessible ports, side cover, power supply unit and accessories
with a disposable cloth and mild detergent or alcohol disinfectant (see list below).
2. Remove any excess disinfectant with a disposable dry cloth.
Agents recommended for surface disinfection and cleaning:
Warm water and mild detergent eg, Teepol
Window cleaner or other premixed surface detergent
Methyl alcohol solution
70% Ethyl alcohol solution
70-90% Isopropanol solution
10% Bleach solution
Isopropyl wipes
CaviCide
Mikrozid
®
Actichlor
Plus
Terralin
.
®
Note: Agents may not be available in all regions.
multipurpose detergent
English 1

Advertisement

loading

Summary of Contents for ResMed AirSense 10

  • Page 1 Note: The standard water tub cannot be disinfected. If contaminated, it must be discarded and replaced with a new water tub. WARNING • ResMed cannot give any assurance that deviations from the procedures listed in this guide, and their effect on the performance of the product, will be acceptable.
  • Page 2: Reprocessing The Air Tubing And Air10 Tubing Elbow

    2. Run the detergent solution through the air tubing repeatedly until no contamination is visible. 3. Thoroughly rinse each component according to the detergent manufacturer's instructions. ResMed has tested the following detergents according to the manufacturer’s instructions: Detergent Water temperature...
  • Page 3: High Level Chemical Disinfection

    High level chemical disinfection Part Validated number of cycles CIDEX OPA Ortho-phthalaldehyde 0.55% at room temperature (approx. 69.8ºF or 21°C) for 12 minutes ® SlimLine ClimateLineAir ClimateLineAir Oxy Standard Air10 tubing elbow 1. Soak the air tubing/Air10 tubing elbow in a commercially available solution of a chemical sterilant. Take care that no air bubbles are trapped inside the air tubing.
  • Page 4: Reprocessing The Water Tub And Air Outlet

    Reprocessing the water tub and air outlet Disassembling The following instructions provide guidance on how to correctly disassemble the cleanable water tub and the air outlet. 1. Remove the water tub from the device, open it and discard any remaining water. 2.
  • Page 5 ResMed has tested the following detergents according to the manufacturer’s instructions: Detergent Water temperature Cleanable water Air outlet Alconox (diluted at 1%) Hot water (approx 140°F or 60°C)   Warm water (approx 113 to 140°F or 45 to 60°C) Room temperature water (approx 70°F or 21°C)
  • Page 6 Reassembling The following instructions provide guidance on how to correctly reassemble the air outlet and the water tub. To reassemble the air outlet 1. Hold the air outlet with the seal pointing to the left and the clip pointing forward. 2.
  • Page 7 Advanced Sterilization Products, Division of Ethicon US, LLC. Mikrozid and Terralin are trademarks of Schülke & Mayr GmbH. Neodisher MediZym is a trademark of Chemische Fabrik Dr Weigert GmbH & Co. Sterrad is a trademark of Johnson & Johnson.Teepol is a trademark of Shell Chemical Co. For patent and other intellectual property information, see Resmed.com/ip.