Manually Modifying The Calibration Parameters; Figure 3-12 Transducer Swap Out Wizard - Emerson Daniel 3814 Maintenance And Troubleshooting Manual

Liquid ultrasonic flow meter
Hide thumbs Also See for Daniel 3814:
Table of Contents

Advertisement

Maintenance and Troubleshooting Manual
3-9000-764 Rev F
3.3.1

Manually modifying the Calibration Parameters

Use the Tools|Transducer Swap-out utility in Daniel MeterLink to easily update parameters
such as path lengths, delay times, and delta times for chord. This is necessary anytime
transducers or housings have to be replaced for a chord or if the meter has performed a cold
start. The wizard steps you through selecting which chords have been changed, which
components in the chords have been changed, as well as recalculating new path lengths based
on the information entered. This utility is only available while connected to a meter. The length
of the meter body is found on the original calibration sheet supplied with the meter.

Figure 3-12 Transducer Swap Out Wizard

Manually modifying the Calibration Parameters
Section 3: Meter maintenance
December 2017
23