Sleep Apnea Therapy User Manual
Device LED
Fault LED
Off
Flashes
14 times*
*
If this fault occurs, check all tubes and hoses to ensure that they are properly connected. If the fault occured during
therapy, make sure the mask is properly fitted and not leaking. Check that all filters are properly installed and not
excessively dirty.
Smartcard fault codes
When the CPAP detects a problem while communicating with the smartcard, it flashes the yellow fault LED to identify the
error. The green device LED is also illuminated during this time to differentiate the fault from the non-smartcard faults
enumerated above. The fault is acknowledged by removing the card.
Device LED
Fault LED
On
Flashes 1
times
On
Flashes 5
times
On
Flashes 6
times
On
Flashes 8
times
Alert codes
Alerts are like faults except they do not reset the processor when they occur (i.e. the CPAP continues as usual).
Device LED
Fault LED
Off
Flashes
15 times
Filter change reminder
Every six months, the green LED flashes as a reminder that the air inlet filter should be changed. The LED only flashes when
the CPAP is in standby mode and the CPAP otherwise still functions as normal. The reminder may be acknowledged in a
manner similar to acknowledging a fault (hold down the power button until the green LED stops flashing). Acknowledging
the reminder resets the reminder timer (i.e. it will return in six months). The reminder should be acknowledged in conjunction
with changing the filter.
Error
Comments
Blower over-temp
The blower thermistor has reported an excessive temperature.
Error
Comments
Communication/
CPAP was unable to establish or maintain communications with the
media error
smartcard. This may be due to a card not seated correctly in the slot, a
bad/dirty slot connector or a defective card. Retry first with the same
card to see if it's a seating issue. Then try a different card to see if it's a
problem in the card or the CPAP.
Could not erase
Probably a defective card but could be seating issue. Retry, if problem
persists, try different card.
Bad prescription
The prescription data on the card is corrupt. Probably caused by
data
defective card but could also indicate problem with software that
created the card.
Wrong prescription
The card, with prescription data on it, is intended for a different serial
data
number/patient than the CPAP being used. The card should be sent
back to the provider.
Error
Comments
Cannot regulate
CPAP repeatedly detected the monitored pressure outside of the
therapy target range. The CPAP attempts to continue to deliver therapy.
The alert is an indication that filters and the mask should be inspected
for cleanliness and proper fit.
Fault, alert and reminder codes
Page 43