This section presents the following topics:
•
Cleaning the interior with drying mode
•
Cleaning the exterior
•
Cleaning components
•
Cleaning the air inlet filter
Cleaning the interior with drying mode
Drying mode should be used daily to clean the Transcend interior.
1.
Set Transcend on a dry and stable surface (cloth surface is not recommended as it may get soiled). Do not
wear Transcend while it is in drying mode.
2.
To use drying mode, Transcend must be plugged in to a power source.
3.
To enable drying mode, press and hold the ramp button (top button) then press the power button (bottom
button).
4.
Low pressure air flow will begin and last for 30 minutes until cleaning session is completed, then it will
return to standby mode. To end the drying mode before the 30 minute session is complete, press the
power button. Transcend will return to standby mode.
Cleaning the exterior
Clean the exterior of your Transcend as needed. Use a damp soft cloth or paper towel with mild dishwashing soap
to wipe clean. Never submerge Transcend in water.
Do not get soap or water in the device.