Operation and Maintenance Manual – Steam sterilizers mod. BABY S Table of contents: 1. INTRODUCTION..................................3 2. T E C H N I C A L S P E C I F I C A T I O N S ......................4 3.
Operation and Maintenance Manual – Steam sterilizers mod. BABY S 1. INTRODUCTION The machine described in this operation and maintenance manual is a steam sterilizer that has been specially designed to sterilize most of the materials and instruments used in dental surgeries, doctors' surgeries and hospitals in general.
7 lt. CAPACITY chamber. Riscaldamento graduale al fine di ottenere una temperatura più precisa e una migliore sterilizzazione. After each cycle, the BABY S autoclave, as required, triggers an anti-condensation system unless the door is opened within a few minutes.
Page 5
Operation and Maintenance Manual – Steam sterilizers mod. BABY S Description Unit of measurement Values Power source voltage rating 230 V.a.c. ~ Power source frequency Machine power rating 1000 Boiler volume Ø 170 x 270 Operating pressure 2.1 - 1.1...
Operation and Maintenance Manual – Steam sterilizers mod. BABY S 3. G E N E R A L R E C O M M E N D A T I O N S 3.1 Regulations for the operator Always wear personal protective equipment in compliance with the directives that establish the current accident-preventing provisions.
4. L O C A T I O N O F T H E C O N T R O L S BABY S feature an elegant casing treated with scratch-proof epoxy powder paint in the RAL 9010 COLOUR. No particular preparations are required to operate the machines. Just fit the plug into a 230 V power socket.
Operation and Maintenance Manual – Steam sterilizers mod. BABY S 5. S T A R T I N G A N D S E T T I N G A T W O R K 5.1 Removing the machine from the packing The machine is consigned packed on a wooden pallet in a cardboard box: make sure that there are no dents and that the wrapping is in optimum conditions.
Page 9
Operation and Maintenance Manual – Steam sterilizers mod. BABY S DRAWING ILLUSTRATING HOW THE MACHINE MUST BE LIFTED FROM THE PACK Rev. 01 del 01-07-2013 - C:\Documenti\Manuali di istruzione\BABY S GB-R01.doc Pag. 9 di 28...
EQUIPPED WITH AN EARTH CONNECTION. Make sure that the mains voltage rating corresponds to the value indicated on the label at the rear of the machine (230 Volts). BABY S must only be filled with DISTILLED WATER to ensure that the electrical and air-operated devices operate efficiently and are long lasting.
Page 11
Operation and Maintenance Manual – Steam sterilizers mod. BABY S DESCRIPTION OF THE REAR PART 1) Pipe can. 2) Water drainpipe used for connecting to the can through "1". The same pipe can be used for draining into the sewage system.
Page 12
Operation and Maintenance Manual – Steam sterilizers mod. BABY S READ OF INFO DISPLAY: 3= Cycle number 15= Sterilization time 5= Dry time 121= Temperature of sterilization °C 1 2 1 4= Cycle number 20= Sterilization time 15= Dry time 121= Temperature of sterilization °C...
Operation and Maintenance Manual – Steam sterilizers mod. BABY S 6. G E N E R A L I N S T R U C T I O N S F O R U S E 6.1 Directions for use OPEN DOOR WITH SWITCH (9) TURNED.
Operation and Maintenance Manual – Steam sterilizers mod. BABY S 7.1 Sterilizing table STERILIZING TIME (Min.) DRYING TIME (Min.) CYCLES FREE STERILIZING TIME (Min.) DRYING TIME (Min.) CYCLES AT 121° 3 UNWRAPPED 4 WRAPPED STERILIZING TIME (Min.) DRYING TIME (Min.) CYCLES AT 134°...
Operation and Maintenance Manual – Steam sterilizers mod. BABY S 7.3 Monthly inspection proposal Fac simile of an inspection form Sterilizer data Brand Model Year of purchase Observations Periodical internal inspection Date Cycle length on Cycle temperature Cycle temperature display 20’...
Operation and Maintenance Manual – Steam sterilizers mod. BABY S 8. MESSAGES DISPLAYED 8.1 Alarm management MESSAGE ERROR TYPE REMEDY NO POWER No mains power Fit the machine plug in a single socket that complies with standards and is without adapters or extensions. Change the fuse and check the causes.
Operation and Maintenance Manual – Steam sterilizers mod. BABY S 9 . M A I N T E N A N C E Carry out the maintenance operations with the machine at a standstill, after having removed the plug from the power socket and after having allowed the components to cool down.
Operation and Maintenance Manual – Steam sterilizers mod. BABY S 9.4 Monthly routine maintenance (important) Lubricate the pins and locking mechanisms. Check the movement of the hinge and door locking pin. Lubricate with Vaseline oil. Make sure that the door is well regulated. If it is excessively slack, it must be adjusted by means of the supplied wrench.
Operation and Maintenance Manual – Steam sterilizers mod. BABY S 9.6 Recommended inspections and replacements (ONLY FOR TECHNICIANS) 1) Make sure that the safety valve functions correctly. Remove the conveyor tube - Carry out the 134° cycle and when the maximum pressure has been reached (2.2 bar), unscrew and retighten the knurled ring nut on the safety valve.
Page 21
Operation and Maintenance Manual – Steam sterilizers mod. BABY S ELECTRICAL SYSTEM Rev. 01 del 01-07-2013 - C:\Documenti\Manuali di istruzione\BABY S GB-R01.doc Pag. 21 di 28...
Page 22
Operation and Maintenance Manual – Steam sterilizers mod. BABY S HYDRAULIC CIRCUIT: Rev. 01 del 01-07-2013 - C:\Documenti\Manuali di istruzione\BABY S GB-R01.doc Pag. 22 di 28...
Page 23
Operation and Maintenance Manual – Steam sterilizers mod. BABY S SPARE PARTS 3023 Pressostat collector 4246 PT100 probe 4343 Solenoid valve n.o. 1067 Bacteriological filter 0336 Foot base 4248 Tank gasket 3136 Y Filter 4191 Rubber foot 3928 Locking pin...
Page 24
Operation and Maintenance Manual – Steam sterilizers mod. BABY S FORM OF RETURN FOR REPARATION OF REVERBERI ARNALDO S.R.L To be filled by the technician AUTOCLAVES Via Don Luigi Sturzo n°6 42021 Barco di Bibbiano or an authorized person Reggio Emilia – ITALY Tel.
Page 25
Operation and Maintenance Manual – Steam sterilizers mod. BABY S It is obligatory for the installer to fill out all parts of this form and to declare that he has complied with all the conditions ensuring that the appliance is correctly installed in compliance with sect. 5.2 of the instruction manual.
Operation and Maintenance Manual – Steam sterilizers mod. BABY S W A R R A N T Y C E R T I F I C A T E This document certifies that the appliance has been correctly manufactured. The appliance will be repaired if it operates in a faulty way during the warranty period owing to manufacturing defects, the manufacturer's judgement over such matters being final.
Page 27
Operation and Maintenance Manual – Steam sterilizers mod. BABY S WARRANTY CERTIFICATE APPLIANCE: The undersigned reseller certifies that the warranty has been examined and certifies that the appliance has been consigned and installed in compliance with the instructions given by the manufacturer on:……………………………………………………………………………………………..
N u m e r o d i s e r i e : … … … … . S e r i a l n u m b e r : … … … … … … La Reverberi s.r.l. dichiara sotto la propria responsabilità che i prodotti elencati sono conformi alle norme e le direttive europee: The company Reverberi s.r.l.