Reverberi BABY S Technician Manual

Reverberi BABY S Technician Manual

Autoclave
Hide thumbs Also See for BABY S:

Advertisement

MEDICAL EQUIPMENT DIVISION
Via Don L. Sturzo,6 – 42021 Barco (RE) (ITALY)
Codice Fiscale e Partita IVA: IT 01363800358
Numero REA: 180051 C.C.I.A.A. di Reggio Emilia
Tel. 0039 (0) 522 875159 Fax 0039 (0) 522 875579
e-mail: reverberisrl@gmail.com
Technician's manual
BABY S
PRATIKA S VACUUM
AUTOCLAVE

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BABY S and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Reverberi BABY S

  • Page 1 Via Don L. Sturzo,6 – 42021 Barco (RE) (ITALY) Codice Fiscale e Partita IVA: IT 01363800358 Numero REA: 180051 C.C.I.A.A. di Reggio Emilia Tel. 0039 (0) 522 875159 Fax 0039 (0) 522 875579 e-mail: reverberisrl@gmail.com Technician’s manual BABY S PRATIKA S VACUUM AUTOCLAVE...
  • Page 2 Technician’s manual – BABY S PRATIKA S steam steriliser ACCESSING TEST MODE TEST 1) Machine off, door open. 2) Press and hold the ENTER and START buttons simultaneously and turn on the machine with the green ON/OFF button. 3) Once the machine is on, release the ENTER and START buttons simultaneously.
  • Page 3 DISPLAY s u p Press the INVIO/ENTER button to activate the vacuum pump and the air outlet solenoid valve. On the BABY S model, only the air outlet solenoid 1 2 3 4 valve is activated. Press the SELECT + button to call up another actuator on the top row of the display.
  • Page 4 Technician’s manual – BABY S PRATIKA S steam steriliser SETTING THE ELECTRONIC BOARD STERILISING TIME CYCLE NUMBER DRYING TIME With the machine off and the door open, press the main ON/OFF power switch. Press the SELECT + button to select cycle 7.
  • Page 5 Technician’s manual – BABY S PRATIKA S steam steriliser ADJUSTING TEMPERATURE Press the + and – buttons to adjust the value (default value Ø). This adjusts the autoclave operating temperature. Set negative values (-1, -2 etc.) to increase the operating temperature, set positive values (+1, +2 etc.) to decrease the...
  • Page 6 Technician’s manual – BABY S PRATIKA S steam steriliser SIGNIFICANCE OF FLASHING DOTS DISPLAYED DURING STERILISATION CYCLE WITH DOOR CLOSED CHAMBER HEATING ATTENTION: PRESSURE SWITCH CALIBRATION FLASHES WHILE WAITING FOR PRESET TEMPERATURE PROCEDURE The following conditions are necessary:  New door seal.

This manual is also suitable for: