Operation and maintenance manual
STEAM STERILIZERS
MODELS
STERILIX 2 PLUS
STERILIX VACUUM PLUS
STERILIX STAR

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the STERILIX 2 PLUS and is the answer not in the manual?

Questions and answers

Summary of Contents for Reverberi STERILIX 2 PLUS

  • Page 1 COMPANY WITH CERTIFIED QUALITY SYSTEM UNI EN ISO 9001 & EN 13485 MEDICAL APPLIANCE DIVISION Cod.3125 Operation and maintenance manual STEAM STERILIZERS MODELS STERILIX 2 PLUS STERILIX VACUUM PLUS STERILIX STAR...
  • Page 2 Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast Reverberi s.r.l. thanks you for having bought one of its products and asks you to read this manual. The manual contains all the information you require in order to use the machine you have purchased.
  • Page 3: Table Of Contents

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast Table of contents: Chapter Description Page Introduction ………………………………………………………………………………………………… Technical specifications …………………………………………………………………………….. General recommendations ………..………………………………………………………………… Regulations for the operator ………………………………………………………………………. Location of the controls …………………………………………………….…………………..……...
  • Page 4: Introduction

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 1 . I N T R O D U C T I O N The machine described in this operation and maintenance manual is a steam sterilizer that has been specially designed to sterilize most of the materials and instruments used in dental surgeries, doctors' surgeries and hospitals in general.
  • Page 5: Technical Specifications

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 2 . T E C H N I C A L S P E C I F I C A T I O N S STERILIX are autoclaves designed and built in compliance with the operators' need for functionality, practical use and safety.
  • Page 6 Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast Description Unit of measurement Values Power source voltage rating 230 V.a.c. ~ Power source frequency Machine power rating 1500 2500 (for Mod.
  • Page 7: General Recommendations

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast . G E N E R A L R E C O M M E N D A T I O N S...
  • Page 8: Location Of The Controls

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 4 . L O C A T I O N O F T H E C O N T R O L S STERILIX feature an elegant casing treated with scratch-proof epoxy powder paint in the RAL 7035 COLOUR.
  • Page 9 Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast Rev. 09 of 18-04-02 - E:\Documenti1\Manuali superati non in uso\Vacuum Star GB-R09.doc Page 9 of 42...
  • Page 10: Starting And Setting At Work

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 5 . S T A R T I N G A N D S E T T I N G A T W O R K Carry out a cycle of dry-run tests using one of our mod.
  • Page 11 Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast DRAWING ILLUSTRATING HOW THE MACHINE MUST BE LIFTED FROM THE PACK Rev. 09 of 18-04-02 - E:\Documenti1\Manuali superati non in uso\Vacuum Star GB-R09.doc...
  • Page 12: Installation And Preliminary Inspections

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 5.2 Installation and preliminary inspections The installer is obliged to make sure that the place in which the sterilizer is installed is fit for that purpose (CEI 62/4).
  • Page 13: Printer Connection

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast DESCRIPTION OF THE REAR PART 1) Can tube to connect to the used water drain tube “3” 2) Screws to fix the casing.
  • Page 14: General Instructions For Use

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 6 . G E N E R A L I N S T R U C T I O N S F O R U S E 6.1 Method of use...
  • Page 15: How To Arrange The Instruments For The Sterilizing Process

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 7 . H O W T O A R R A N G E T H E I N S T R U M E N T S F O R...
  • Page 16: Program Selection

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 8 . P R O G R A M S E L E C T I O N 8.1 121ºC program Ideal for handpieces, other types of rotating instruments, various types of packaged and loose instruments, plastic and liquid materials.
  • Page 17: 134ºc Program

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 8.2 134ºC program Ideal for packaged and loose instruments and for instruments without mechanisms. OPEN DOOR WITH SWITCH (10) TURNED OFF IN “0” POSITION (IMPORTANT) Place the trays containing the instruments to be sterilized into the machine.
  • Page 18: 121°/134° Fractioned Program (For Mod. Star And Star Fast Only)

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 8.3 121°/134° fractionated program (for mod. STAR and STAR FAST only) Optimum for packeted and loose instruments, hollow items and boxes OPEN DOOR WITH SWITCH (10) TURNED OFF IN “0”...
  • Page 19: Bowie And Dick Test

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 8.5 Bowie & Dick Test (for mod. Sterilix Star, Star Fast and V. Plus only) OPEN DOOR WITH SWITCH (10) TURNED OFF IN “0” POSITION (IMPORTANT) Insert the trays containing the packet in the center to conduct the test for steam autoclaves with vacuum pump.
  • Page 20: Sterilizing Chart

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 8.6 Sterilizing table Rev. 09 of 18-04-02 - E:\Documenti1\Manuali superati non in uso\Vacuum Star GB-R09.doc Page 20 of 42...
  • Page 21: Notes And Warnings

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 8.7 Notes and warnings 1- Strictly comply with the manufacturer's instructions when sterilizing materials. 2- To preserve packaged and loose sterilized materials, the trays must be stored in special cabinets fitted with ultraviolet lamps and designed for this specific use.
  • Page 22: Cycle Charts

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 8.8 Cycle charts STANDARD CYCLE (mod. STERILIX 2 PLUS only) 1) Balancing time 2) Sterilizing and T° upkeep 3) Drying 4) Upkeep...
  • Page 23: Monthly Inspection Proposal

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 8.9 Monthly inspection proposal Fac simile of an inspection form Sterilizer data Brand Model Year of purchase Observations Periodical internal inspection...
  • Page 24: Example Of Correct Cycle Prints

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 8.10 Examples of correct cycle prints Fractioned 134° DEDUCTION 134° DEDUCTION 121° DEDUCTION (only STAR and STAR FAST) (only VACUUM PLUS, STAR and...
  • Page 25: Useful Information

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 9 . U S E F U L I N F O R M A T I O N A – There is a DEDUCTION (1) key on the control panel that allows users to shorten the sterilizing time (from 1 to 12 minutes) if required, while maintaining the length of the drying cycle unchanged.
  • Page 26: Key To Displayed Messages

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 1 0 . K E Y T O D I S P L A Y E D M E S S A G E S Indicates a block status.
  • Page 27 Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 29 - ElS ( For technicians only ): Draining solenoid valve shut. (15) 30 - PoF Machine status message. Pressure less than 0.8 bar. (1) 31 - Pon Machine status message.
  • Page 28: Troubleshooting: Solutions / Remedies

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 10.1 Troubleshooting: Solutions / Remedies 01- tPt Repeat the sterilizing process or contact the after-sales service center. 02- rEt Wait until the power supply returns. Check the socket and replace it if necessary.
  • Page 29: Maintenance

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 12- dPr Too much moisture in the chamber before the sterilizing cycle starts: after having opened the door, deactivate the alarm by pressing the selected program for 3 sec., then repeat the cycle.
  • Page 30: Daily Routine Maintenance

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 11.2 Daily routine maintenance Clear the door seal, door window and generally clean the outer and inner surfaces of the machine.
  • Page 31: Periodic Extraordinary Maintenance

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 11.5 Periodic extraordinary maintenance To allow the autoclave to operate in a regular way, the manufacturer advises users to carry out a functional test to ensure that the temperature and pressure parameters of the machine comply with the standards required for a correct sterilizing process.
  • Page 32: Faults Without Messages

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast 1 2 . F A U L T S W I T H O U T M E S S A G E S 1) The machine fails to come on when the main switch is turned to position 1.
  • Page 33: Sterilix 2 Plus Electrical System

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast STERILIX 2 PLUS ELECTRICAL SYSTEM: Rev. 09 of 18-04-02 - E:\Documenti1\Manuali superati non in uso\Vacuum Star GB-R09.doc Page 33 of 42...
  • Page 34: Sterilix Vacuum Plus/Sterilix Star Electrical System

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast STERILIX VACUUM PLUS / STERILIX STAR ELECTRICAL SYSTEM: Rev. 09 of 18-04-02 - E:\Documenti1\Manuali superati non in uso\Vacuum Star GB-R09.doc Page 34 of 42...
  • Page 35: Sterilix Star Fast Electrical System

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast STERILIX STERILIX STAR FAST ELECTRICAL SYSTEM: Rev. 09 of 18-04-02 - E:\Documenti1\Manuali superati non in uso\Vacuum Star GB-R09.doc Page 35 of 42...
  • Page 36: Sterilix 2 Plus Hydraulic Circuit

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast STERILIX 2 PLUS HYDRAULIC CIRCUIT: Rev. 09 of 18-04-02 - E:\Documenti1\Manuali superati non in uso\Vacuum Star GB-R09.doc Page 36 of 42...
  • Page 37: Sterilix Vacuum Plus / Star / Star Fast Hydraulic Circuit

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast VACUUM PLUS / STAR / STAR FAST HYDRAULIC CIRCUIT: Rev. 09 of 18-04-02 - E:\Documenti1\Manuali superati non in uso\Vacuum Star GB-R09.doc...
  • Page 38: Form To Use When Appliances Are Returned For Repairs

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast FORM TO FILL OUT WHEN To be filled out by the: APPLIANCES ARE Via Don Luigi Sturzo n°6 42020 BARCO ( Reggio Emilia ) ITALY...
  • Page 39 Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast It is obligatory for the installer to fill out all parts of this form and to declare that he has complied with all the conditions ensuring that the appliance is correctly installed in compliance with sect.
  • Page 40: Warranty Certificate

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast WARRANTY CERTIFICATE This document certifies that the appliance has been correctly manufactured. The appliance will be repaired if it operates in a faulty way during the warranty period owing to manufacturing defects, the manufacturer's judgement over such matters being final.
  • Page 41 Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast WARRANTY CERTIFICATE and stamped APPLIANCE: The undersigned reseller certifies that the warranty has been examined and certifies that the appliance has been consigned and installed in compliance with the instructions given by the manufacturer on:……………………………………………………………………………………………..
  • Page 42: Declarations Of Conformity

    Operation and Maintenance Manual – Steam sterilizers mod. Sterilix 2 Plus & Vacuum Plus & Star & Star Fast Via Don Luigi Sturzo nº 6 42020 BARCO (Reggio Emilia) Tel. +39 – 0522 - 875159 / 875195 Fax +39 – 0522 - 875736 / 875579...