Vtex1629_Self_Cal_Clear - VTI Instruments EX1629 User Manual

48-channel strain gage instrument
Table of Contents

Advertisement

F
P
UNCTION
ROTOTYPE
ViStatus vtex1629_self_cal_clear (ViSession vi);
F
P
UNCTION
ARAMETERS
vi = contains a session handle to the instrument. This handle is obtained by the function and remains valid until the
session is closed.
D
I
R
V
ATA
TEM
ESET
ALUE
Not applicable to this function.
D
ESCRIPTION
This function clears the current self-calibration data. This operation clears the volatile data, but does not affect any
self-calibration data that is stored in nonvolatile memory. After this function, the effective calibration will be the full
calibration (factory calibration, also known as the annual calibration) plus, if it exists, the nonvolatile self-
calibration. If a self-calibration is performed, and non-volatile self-calibration data exists, executing this function
effectively reverts to the non-volatile self-calibration data. With no non-volatile self-calibration data, this effectively
reverts to full (factory/annual) calibration data.
E
XAMPLE
ViSession instrumentHandle;
ViStatus status;
...
status = vtex1629_self_cal_clear (instrumentHandle);
EX1629 Command Set

vtex1629_self_cal_clear

www.vtiinstruments.com
219