 Problem :
The message: 'NO SCIP' is shown in the program bar of the SCIP program. It stays on the bar
even if the SCIP tester is connected to the Archer
 Solution :
 Check that the SCIP tester's On-Off switch is at On and that the LED is on.
 Verify that the SCIP tester's battery is powered enough and not within the critical threshold
limit.
 In Cable mode, verify that the cable is plugged correctly into the SCIP tester and into the
Archer
2
pocket PC.
 In Bluetooth mode, check if a Bluetooth Partnership has been established between your
SCIP Tester and your Archer
 Problem:
In Bluetooth mode, a COM Error message appears:
Instrumentation GDD Inc.
2
2
PDA (see Section 11.2 – Bluetooth Partnership).
2018-09-18
PDA.