Troubleshooting; Problems - GDD Instrumentation TDLV Instruction Manual

Sample core ip tester
Table of Contents

Advertisement

11 Troubleshooting

11.1 Problems

This section explains some problems that could occur while using the SCIP tester as well as some
proposed solutions.
For any issues regarding the Archer
program, please refer to Archer
CD/USB stick, sent by GDD.
 Problem :
The SCIP tester does not turn on when the On-Off switch is at 'On'.
 Solution :
 If the SCIP tester's battery power rating is below the critical threshold, the SCIP tester will
not come on line. (See the Power section for more details.)
 Problem :
The message "Battery Error" appears on the main screen of the SCIP program on the Archer
PDA.
 Solution :
 A problem has occurred while charging the SCIP battery: over voltage, charging under 0°C
or over 45°C, the charging time was too long, the battery is defective, etc.
 Try to disconnect and re-connect the AC charger.