    Contents Fixing the Weighing ---------------------------------------------------------------------- Protection Device ---------------------------------------------------------------------- Disassembly Steps Precautions ---------------------------------------------------------------------- Descriptions of Keys ---------------------------------------------------------------------- Functional Operation ---------------------------------------------------------------------- of Keys Panel Illustration ---------------------------------------------------------------------- Single Point ---------------------------------------------------------------------- Calibration of Weight Linearity Calibration ---------------------------------------------------------------------- of Weight Parameter Setting ---------------------------------------------------------------------- Parameter Functional ---------------------------------------------------------------------- Description Units Conversion...
  • Page 2: Fixing The Weighing

    Weighing Pan Fixing Steps 1. When installing the 2. Unscrew and remove the 3. After removing the shipping weighing pan, tighten the shipping protection screw on protection screw on the side of fixing screw in the centre of the side of the scale. the scale, fasten the shipping the plastic weighing pan.
  • Page 3: Descriptions Of Keys

    Descriptions of Keys ON/OFF: The on/off key for switching on/off the equipment. The zero key for resetting the equipment, functional only when weight is within 4% of the weighing ZERO: capacity. TARE: The tare key for deducting the weight of container on the weighing pan. MODE: The switching key to select weighing, counting and percentage modes in turn.
  • Page 4: Panel Illustration

    Panel Illustration Level Stable indicator Warming up indicator Tare key Sampling/Backlight key Zero key Function selection key Weight display window Unit conversion key Single Point Calibration of Weight Step 1: Power off while holding down 「MODE」key, and then press 「ON/OFF」to switch on the scale.
  • Page 5: Linearity Calibration

    CAL-S The calibration can be done at the point of 1/3 load, 2/3 load or full load. Choose the one you need and place the corresponding weight onto the pan. If place 1/3 of load onto the pan, the screen will show: If place 2/3 of load onto the pan, the screen will show: If place full load onto the pan, the...
  • Page 6: Parameter Setting

    Step 8: After countdown, the linearity calibration is completed and the scale returns to the normal weighing mode. Parameter Setting Hold down 「MODE」 key, then press 「ON/OFF」 key to enter the setting mode. The screen will show Count 「MODE」key: To select functions; 「UNIT」key: To enter or exit function setting selection;...
  • Page 7: Parameter Functional

    ↓ Pr Fu External device PC/ SH-24 Item 10, Page 7 ↓ diV-Z Display range Zero0/1/2/3/4/5 Item 11, Page 7 When the LCD display shows Count, press the「UNIT」key, the current internal value will be displayed. Press 「UNIT」to return to LCD display Count. unit , press 「...
  • Page 8: Units Conversion

    X represents the backlight setting. Press 「UNIT」 to select The LCD display XXXX LigH 。 the desired backlight setting and press「MODE」to return to LCD display 『 』 onoFF- means auto-backlight, oFF-means backlight off, ON-means backlight on When the LCD display shows Print , press「UNIT」...
  • Page 9: Cross-Reference

    Cross-reference Table of Full Loads 1500 3000 Gram 150X0.005 300×0.01 600×0.02 1500×0.05 3000×0.1 Carat 750.34X0.02 1500.70X0.05 3001.36X0.1 7503.4X0.2 15007.0X0.5 Pound(Avoir) 0.33080X0.00002 0.66160X0.00005 1.323X0.0001 3.3080X0.0002 6.6160X0.0005 Ounce(Avoir) 5.2926X0.0002 10.5855X0.0005 21.171X0.001 52.926X0.002 105.855X0.005 Dram(Avoir) 84.685X0.005 169.37X0.01 338.72X0.02 846.85X0.05 1693.7X0.1 Grain 2315.5X0.1 4631.0X0.2 9262.0X0.5 23155X1 46310X2...
  • Page 10 BIT FORMAT: START BIT DATA BIT STOP BIT DATA FORMAT: S/U T LF CR WEIGHT UNIT ST:STABLE UT:UNSTABLE GW:GROSS WEIGHT SP:SPACE Examples: ST GW + 150.00 g 80.00 ct ………….. UT NW - PC Program: 10 OPEN〝COM1:9600,N,8,1,RS,DS,LF〞AS #1 (BASICA) 20 INPUT #1,A$ 30 PRINT A$ 40 GOTO 20 50 END...