26. Decontamination / Cleaning

CAUTION
Do not immerse the battery, helmet and fan module into water or other liquid. This will cause irreparable damage to
the helmet.
Do not use solvent or alcohol to clean the helmet. Isopropyl alcohol may be used to clean the Helmet. However,
repeated long term use of isopropyl alcohol may deface the Helmet.
Do not subject helmet to any sterilization cycles.
Do not use organic solvents or strong oxidizing agents to clean the helmet.
The air channels should never need cleaning. If they do, the Filters are not being maintained properly or replaced at
the appropriate intervals.
If other cleaning agents are to be used, it is recommended to test their use on a small section of one DLC Lens and/
or a small section of the Helmet Liner to determine short and long term side effects.
It is not recommended to disconnect the Power Cord from the Helmet. The Power Cord should be decontaminated
and cleaned as part of the Helmet.
Decontaminating
Supplies Needed:
● Decontaminating wipe
● Decontaminating Agent: Alcohol,
bleach, or quaternary ammonia.
Procedure:
1. Inspect the system and perform any
assembly/disassembly instructions
necessary for disposable items
and for all components that have
become worn or damaged.
NOTE
If desired, replace the DLC, Helmet Liner, Filter Cartridge, or Filter Cover Cap by following their assembly and
disassembly procedures.
Replace the Front Headband Comfort Strip with a new one.
The rear Closed Cell Foam comfort pad may be cleaned for reuse by wiping down the outer surfaces with a
decontaminating wipe.
Cleaning
Supplies Needed:
● Clean Damp Cloth
● Cleaning Agent: Mild application of
skin friendly soap.
Procedure:
1. Use a damp cloth with cleaning
agent to clean all outer and inner
exposed surfaces.
NOTE
If desirable, replace the damaged or soiled Front Headband Comfort Strip.
The rear Closed Cell Foam comfort strip may be cleaned for reuse.
Page 103
P/N 03521015 Rev Q
Frequency:
● Wipe between uses and between
different users wearing the system.
2. Apply a suitable wipe with a
decontaminating agent over all
outside reachable surfaces, and
then over all inside surfaces.
Frequency:
● Wipe between uses and between
different users wearing the system.
2. Let air dry.
Accomplishes:
● Reduces cross contamination.
● Extends useful life.
● Improves hygiene.
3. Let air dry and re-assemble or
place in storage.
Accomplishes:
● Reduces cross contamination.
● Extends useful life.
● Improves hygiene.