Installation

Network requirements

ATS installation

3-6
Category 5 100Base-T network drop cables with RJ-45 connectors are required for
the following Unity Network connections.
Unity Network IX network
Unity Network MC network
Unity Network RX network
WARNING
LOSS OF ALARMS/DATA — The ATS running ApexPro software
v4 is compatible with CIC Pro centers (BCM, Nightshade, and
Omnitech) running ApexPro v3.9 and above. Sharing of the same
care unit names between ATS and CIC Pro centers having
incompatible ApexPro software versions can result in lost or
corrupted telemetry alarm defaults data and loss of audible alarms.
WARNING
LOSS OF FUNCTION/EQUIPMENT FAILURE — Only external
devices specifically designed to be connected to the ATS, or
approved by GE for use with the ATS, should be connected, as
specified in this manual or as otherwise specified by the
manufacturer. Loss of telemetry function or equipment failure could
result if this instruction is not followed.
WARNING
SHOCK HAZARD — Connect only approved devices to the ATS
according to the instructions in this manual. A shock hazard may
exist if external devices are connected differently from described in
this manual, or as directed by the manufacturer.
WARNING
EQUIPMENT FAILURE — External equipment must be connected
to the ATS only by qualified biomedical engineering personnel. Loss
of telemetry function or equipment failure could result if this
instruction is not followed.
WARNING
LOSS OF FUNCTION — All external cabling used with the ATS
must be routed so as not to interfere with access or operation of the
system.
ApexPro™ Telemetry
2001989-203C