GE ApexPro Service Manual page 137

Telemetry server
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

2001989-203C
Top hard drive
1.
Record the serial number of the new hard drive you are installing.
2.
Complete the preparation procedures.
7-3.
3.
Remove the chassis cover.
4.
Remove the drive bay.
5.
Using a phillips screwdriver, remove the 4 anchor screws securing the CD/hard
drive carrier in the drive bay on both sides.
6.
Slide the CD/hard drive carrier out of the front of the drive bay.
7.
Using a phillips screwdriver, remove the 4 screws securing the top hard drive in
the CD/hard drive carrier on both sides.
8.
Slide the top hard drive out of the back of the CD/hard drive carrier.
9.
Using a phillips screwdriver, remove the screw securing the temperature sensor
to the back left of the hard drive.
ApexPro™ Telemetry
Parts lists, drawings and replacement
See Field replacement preparation on page
See Chassis cover on page 7-6.
See Drive bay on page 7-16.
018A
160A
063A
069A
7-19