Log Onto The Webmin Service Interface - GE ApexPro Service Manual

Telemetry server
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

Log onto the Webmin service interface

Log onto the local Webmin service interface
Log onto the remote Webmin service interface
2001989-203C
You can access the Webmin service interface locally from the ATS, or remotely from
a service laptop connected to the Unity Network IX network.
NOTE
If you are remotely logged onto ATS using VNC and using a command window,
you must exit out of the command window before exiting the VNC.
NOTE
You may need to start a VNC connection and then access Webmin to perform
certain functions.
Complete the following procedure to log onto the Webmin service interface using the
ATS's Browser function.
1.
From the Favorites menu, select LocalWebmin.
2.
On the Login to Webmin screen, make the following entries:
a.
In the Username field, type biomed.
b.
In the Password field, type Change<space>Me.
3.
Select Login.
Complete the following procedures to log on to the Webmin service interface using a
remote service laptop connected to the Unity Network IX network.
Intranet network configuration
The intranet computer (usually an FE laptop) must be TCP/IP configured for
contacting the ATS before you can connect to Webmin:
The intranet computer IP address scheme must be the same as the ATS.
Optional subnet mask and default gateway settings on the intranet computer must
match the ATS.
If necessary, refer to the ATS IP address labels or the site survey documentation for
ATS intranet domain settings.
Intranet computer LAN configuration
It may be necessary to change the Internet Explorer LAN settings on the intranet
computer to connect to Webmin.
Complete the following procedure to change the Internet Explorer LAN settings:
1.
Start the Microsoft Internet Explorer application.
2.
From the Internet Explorer application, select Tools > Internet Options.
ApexPro™ Telemetry
Service interfaces
4-7

Advertisement

Table of Contents
loading

Table of Contents