Log Into Ats Via Vnc; Verify That The Rebuild Procedure Completed Successfully; Note; Top Hard Drive - GE ApexPro Service Manual

Telemetry server
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

Parts lists, drawings and replacement

Log into ATS via VNC

Verify that the rebuild procedure completed successfully

7-20
10. Reverse the steps in this procedure to install the top hard drive.

NOTE

Plan to have alternative monitoring for patients admitted to the tower that is
being repaired by either moving them to another ATS using the Bed
Duplicator application that can be run from the PTS version directory or by
asking the hospital staff for bedside monitoring of the telemetry patients
until the mirroring procedure is complete.
1.
After completing all the steps to replace the defective hard drive, power on the
ATS.
NOTE
The PromiseFastTrack TX2+ application starts to replicate the good
hard drive to the replaced hard drive.
The ATS beeps three times when startup is complete.
2.
Connect the FE laptop to the ATS.
3.
Log into ATS through VNC.
1.
On the ATS desktop, double-click the Promise Array Management (PAM)
icon.
2.
Right-click on the drive icon and select Login.
3.
From the Login dialog box:
a.
In the User name field, type administrator.
b.
Leave the Password field blank.
c.
Select OK.
4.
Navigate to Logical Disk View in the left-hand tree view, and expand that entry.
ApexPro™ Telemetry
See Bed Duplicator on page 4-13.
See Access the ATS desktop on page 4-5.
070A
255A
2001989-203C