Specifications
Specifications are subject to change without notice.
Specification
Dimensions (W x D x H)
Enclosure
Weight
Pollution degree
Over voltage category
Protection class
Power requirements
Fuse
Operating temperature
Storage temperature
Humidity
Altitude
Measurement device
connections
Network connections
Relay card (optional)
Analog output card (optional)
Certification
Warranty
1
The AC power outlets only supply power when the instrument has the optional 100 to 240 VAC
power supply.
Details
Controller: 315 x 120 x 242 mm (12.28 x 4.68 x 9.5 in.)
USB box: 79.5 x 55.1 x 159.5 mm (3.13 x 2.17 x 6.28 in.)
Controller: Metal with corrosion-resistant surface, NEMA 4X/IP65 rating
USB box: ABS/polycarbonate, NEMA 4X/IP65 rating
Approximately 5 kg (11 lb). Weight varies by model.
2
II
I
100 to 240 VAC ± 10 VAC, 50/60 Hz, 1000 VA maximum
F1 and F2: M 3.5 A L, 250 V or T 3.15 A L, 250 V; F3 and F4: T 8 A H, 250 V
-20 to 55 °C (-4 to 131 °F)
-20 to 70 °C (-4 to 158 °F)
95% relative humidity, non-condensing
2000 m (6561 ft)
Two, four or six device connectors and two AC power outlets
Two Ethernet connectors (10/100 Mbps), switch function, M12 female D-coding
connector
One USB connector in a USB box
Four relays on each relay card, change over contacts (SPDT)
Maximum switching voltage: 250 VAC, 125 VDC
Maximum switching current: 5 A
Note: Make sure to install the 5 A external breaker.
Maximum switching power: 1500 VA, 250 VAC; 625 W, 125 VDC
2
Wire gauge: 1.5 mm
(15 AWG) maximum
Four 4-20 mA analog outputs on each analog output card, 500 Ω maximum
2
Wire gauge: 1.5 mm
(15 AWG) maximum
The manufacturer recommends that signal cables with a shield are used.
cTUVus compliant, CE compliant, DIN EN 61326 surge protection