Tektronix AWG5200 Series Installation And Safety Instructions page 27

Arbitrary waveform generators
Hide thumbs Also See for AWG5200 Series:
Table of Contents

Advertisement

3.
Click Diagnostics & Calibration.
4.
In the Diagnostics and Calibration screen, click Calibration.
All calibration routines are selected by default and cannot be deselected.
5.
Click Start.
The Start button changes to Abort once the calibration has started. Clicking Abort stops the calibration process and
restores all values to their previous state. All calibration items should indicate Pass. If not, contact your local Tektronix
service personnel.
AWG5200 Series Installation and Safety Instructions
Install the instrument
11