Safety Information - Fluke Biomedical ESA620 Installation And Getting Started Manual

Electrical safety analyzer
Table of Contents

Advertisement

Safety Information

In this manual, a Warning identifies conditions and
procedures that are dangerous to the user. A Caution
identifies conditions and procedures that can cause
damage to the Product or the equipment under test.
 Warning
To prevent possible electrical shock or
personal injury:
Use this Analyzer only in the manner
specified by the manufacturer or the
protection provided may be impaired.
Read the Users Manual before operating
the Analyzer.
Do not connect the Analyzer to a patient
or equipment connected to a patient. The
Analyzer is intended for equipment
evaluation only and should never be
used in diagnostics, treatment or in any
other capacity where the Analyzer would
come in contact with a patient.
Do not use the product in wet locations,
around explosive gases or dust.
Electrical Safety Analyzer
Inspect the Analyzer before using it. Do
not use the Analyzer if abnormal
conditions of any sort are noted (such as
a faulty display, broken case, etc.)
Inspect the test leads for damaged
insulation or exposed metal. Check test
lead continuity. Replace damaged leads
before using the Analyzer.
When testing, always be sure to keep
your fingers behind the safety barriers
on the test leads.
Never open the Analyzer's case.
Dangerous voltages are present. There
are no user replaceable parts in the
Analyzer.
Have the Analyzer serviced only by
qualified personnel.
Do not use the 15-20 A adapter to power
devices rated in excess of 15 A. Doing so
may overload the installation.
Safety Information
3