Fluke Biomedical ESA620 Installation And Getting Started Manual page 31

Electrical safety analyzer
Table of Contents

Advertisement

Accuracy
Two Terminal Mode
Test current >200 mA ac into 500 mΩ ........ ±(2 % of reading + 0.015 Ω) for 0.0 to 2.0 Ω
Test current 1-16 A ac ................................. ±(2 % of reading + 0.015 Ω) for 0.0 to 0.2 Ω
Four Terminal Mode
Test current >200 mA ac into 500 mΩ ........ ±(2 % of reading + 0.005 Ω) for 0.0 to 2.0 Ω
Test current 1-16 A ac ................................. ±(2 % of reading + 0.005 Ω) for 0.0 to 0.2 Ω
Additional error caused by series inductance
Resistance
0 µH
0.000 Ω
0.000 Ω
0.020 Ω
0.000 Ω
0.040 Ω
0.000 Ω
0.060 Ω
0.000 Ω
0.080 Ω
0.000 Ω
0.100 Ω
0.000 Ω
>0.100 Ω
0.000 Ω
±(5 % of reading + 0.015 Ω) for 0.2 to 2.0 Ω
±(5 % of reading + 0.005 Ω) for 0.2 to 2.0 Ω
Series Inductance
100 µH
200 µH
400 µH
0.030 Ω
0.040 Ω
0.050 Ω
0.025 Ω
0.030 Ω
0.040 Ω
0.020 Ω
0.025 Ω
0.030 Ω
0.015 Ω
0.020 Ω
0.025 Ω
0.010 Ω
0.015 Ω
0.020 Ω
0.010 Ω
0.010 Ω
0.015 Ω
0.010 Ω
0.010 Ω
0.010 Ω
Electrical Safety Analyzer