Mains on applied part test voltage ...................... 110 % ±5 % of Mains, current limited to 7.5 mA ±25 % @ 230 V for IEC 60601
For Alternative and Direct applied parts leakage tests, the leakage values are compensated for nominal mains as
per 62353. Therefore, the accuracy specified for other leakages is not applicable. The actual leakage readings
given during these tests will be higher.
Differential leakage
Ranges ................................................................ 50 to 199 µA
Accuracy.............................................................. ±10 % of reading ±(2 counts or 20 µA, whichever is greater)
Insulation resistance
Ranges ................................................................ 0.5 to 20 MΩ
Accuracy
20 MΩ Range .................................................. ±(2 % of reading + 2 counts)
100 MΩ Range ................................................ ±(7.5 % of reading + 2 counts)
Source test voltage ............................................. 500 V dc (+20 %, -0 %) 1.5 mA short-circuit current or 250 V dc selectable
Maximum load capacitance ................................. 1 µF
100 % ±5 % of Mains for AAMI, current limited to 1 mA ±25 % @ 115 V per AAMI
100 % ±5 % of Mains for 62353 current limited to 3.5 mA ±25 % @ 230 V per 62353
Note
200 to 2000 µA
2.00 to 20.00 mA
20 to 100 MΩ
Electrical Safety Analyzer
Detailed Specifications
21