Poor Linearity
OTHER PROBLEMS:
Cannot calibrate
Calibration weight motor
does not stop
RS-232 not working
Display dark, keys beep
*To be carried out by authorised technicians only.
Find Quality Products Online at:
© 2014
Usually a
Re-check repeatability
mechanical
*Inspection of the flexures for damage or
problem
loose hardware may be required
*Use the Linearity Function in the service
menu to reset linearity
*A problem in the analogue circuit board or
power supplies can cause poor linearity.
Electronic
Make sure all mechanical problems have
Problems
been eliminated first
Zero shifted more
*Check all flexures for damage
than allowed
*Reset dealer calibration
*Verify linearity and repeatability
Calibration
*The balance may be unstable. Verify
timeout
stability as above.
*Check the cables to the motor, try
plugging the balance into the power again
*Look for friction in the calibration weight
movement
*Check the opto-coupler that controls the
motor position.
Doesn't print
Check parameters match the device
connected
Verify cable is correct
*RS-232 circuits damaged
Display contrast
*Check the cables to the display
poor
*Replace the display which could be
Cable unplugged
damaged
or damaged