Specifications; Electrical Specifications; Table 2: Operating Mode; Table 3: Waveforms - Tektronix AFG1022 Technical Reference

Arbitrary function generator specifications and performance verification
Hide thumbs Also See for AFG1022:
Table of Contents

Advertisement

Specifications

Electrical specifications

Table 2: Operating mode

Characteristic
Run mode

Table 3: Waveforms

Characteristic
Standard
Numbers of channel
Arbitrary waveform
Waveform length
AFG1022
AFG1062
Sampling rate
AFG1022
AFG1062
Resolution
Number of non-volatile
waveform memories
Memory size
AFG1000 Series Specifications and Performance Verification
These specifications apply to the AFG1000 Series Arbitrary Function Generator.
All specifications are guaranteed unless labeled "typical". Typical specifications
are provided for your convenience but are not guaranteed.
Specifications that are check marked with the
indirectly) in the Performance tests section.
All specifications apply to the arbitrary function generator unless noted otherwise.
These specifications are valid under the following conditions:
The instrument must have been calibrated/adjusted at an ambient temperature
between +20 °C and +30 °C.
The instrument must be operating at an ambient temperature between 0 °C
and +40 °C.
The instrument must have had a warm-up period of at least 30 minutes.
The instrument must be in an environment with temperature, altitude and
humidity within the operating limits described in these specifications.
Description
Continuous, Modulation, Sweep, and Burst
Description
Sine, Square, Ramp, Pulse, Noise
2
Exponential rise, Exponential fall, Sin(x)/x, Staircase, DC, etc. 45 built-in waveforms, User Defined
2 to 8,192
2 to 1M
125 MS/s
300 MS/s
14 bits
256
64 Mbyte for waveform storage
symbol are checked directly (or
3